| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.052 | 1.944 |
| 5.00 | 1.720 | 1.612 |
| 2.50 | 1.297 | 1.189 |
| 1.25 | 0.752 | 0.644 |
| 0.63 | 0.541 | 0.433 |
| 0.32 | 0.306 | 0.198 |
| 0.16 | 0.190 | 0.082 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-93% | 85% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 90-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-99% | 98-106% | 86-99% | 78-95% |
| EDTA plasma(n=5) | 79-91% | 89-103% | 82-95% | 86-99% |
| Heparin plasma(n=5) | 92-105% | 85-101% | 85-94% | 87-96% |

關注微信公眾號