| 48T | ¥1960.00 | 
| 96T | ¥2800.00 | 
| 96T*5 | ¥11900.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.063 | 1.974 | 
| 5.00 | 1.521 | 1.432 | 
| 2.50 | 1.243 | 1.154 | 
| 1.25 | 0.949 | 0.860 | 
| 0.63 | 0.465 | 0.376 | 
| 0.31 | 0.371 | 0.282 | 
| 0.16 | 0.250 | 0.161 | 
| 0.00 | 0.089 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 83-104% | 94% | 
| EDTA plasma(n=5) | 88-102% | 95% | 
| Heparin plasma(n=5) | 92-102% | 97% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-109% | 83-106% | 87-98% | 83-93% | 
| EDTA plasma(n=5) | 90-99% | 86-118% | 81-95% | 91-120% | 
| Heparin plasma(n=5) | 84-117% | 81-96% | 81-97% | 94-115% | 

            關注微信公眾號