| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.257 | 2.167 |
| 5.00 | 1.541 | 1.451 |
| 2.50 | 1.177 | 1.087 |
| 1.25 | 0.886 | 0.796 |
| 0.63 | 0.567 | 0.477 |
| 0.31 | 0.315 | 0.225 |
| 0.16 | 0.253 | 0.163 |
| 0.00 | 0.090 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-114% | 100% |
| EDTA plasma(n=5) | 94-126% | 110% |
| Heparin plasma(n=5) | 92-112% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-98% | 85-108% | 93-114% | 82-120% |
| EDTA plasma(n=5) | 92-104% | 83-97% | 87-98% | 94-112% |
| Heparin plasma(n=5) | 82-115% | 95-116% | 86-115% | 84-116% |

關注微信公眾號